Тип публикации: статья из журнала
Год издания: 2009
Идентификатор DOI: 10.1134/S1061934809040145
Ключевые слова: Anion concentrations, Electro-osmotic flows, Internal standard methods, Migration time, Peak areas, Spectrophotometric detections, Capillary electrophoresis, Electrochemistry, Electrophoretic mobility, Fluid dynamics, Hydrodynamics, Ions, Polarization, Spectrophotometers, Spectrophotometry, Electroosmosis
Аннотация: A version of capillary electrophoresis with indirect spectrophotometric detection and the hydrodynamic suppression of electroosmotic flow is studied. It is shown that, to improve the reliability of ion identification, one should calculate electrophoretic mobilities of ions or migration times corrected with regard to the electroosmotic flow rate. Correlations between electrophoretic peak areas of ions and their electrophoretic mobilities are derived. In the studied version of capillary electrophoresis, the accuracy of measuring anion concentrations can be improved using the internal standard method.
Издание
Журнал: JOURNAL OF ANALYTICAL CHEMISTRY
Выпуск журнала: Vol. 64, Is. 4
Номера страниц: 398-403
ISSN журнала: 10619348
Место издания: NEW YORK
Издатель: MAIK NAUKA/INTERPERIODICA/SPRINGER
Персоны
- Kalyakin S.N. (Institute of Chemistry and Chemical Technology,Siberian Branch,Russian Academy of Sciences)
- Sursyakova V.V. (Institute of Chemistry and Chemical Technology,Siberian Branch,Russian Academy of Sciences)
- Burmakina G.V. (Institute of Chemistry and Chemical Technology,Siberian Branch,Russian Academy of Sciences)
- Rubailo A.I. (Siberian Federal University)
Вхождение в базы данных
Информация о публикациях загружается с сайта службы поддержки публикационной активности СФУ. Сообщите, если заметили неточности.